Test indicator calibration
The fast, economical and accurate way to calibrate a quantity of
test indicators is to invest in a Dial Indicator Calibrator with the
Test Indicator attachment. These mechanical devices are available in
inch or metric models from several manufacturers. They are in effect
a micrometer head with a large 3.5" diameter, .00005"
accuracy and 0-1" range. The test indicator is positioned above
the spindle using the test indicator attachment. The micrometer head
is rotated and readings are compared. It will be necessary to have
this unit regularly calibrated by a calibration lab to maintain
traceability. Ideally, readings should be taken at every numeral
printed on the test indicator dial, or as your quality manual
requires.
If you need to calibrate large quantities of analog and/or digital
indicators you may want to invest in the electronic i-Checker which
is hooked up to a computer system and generates inspection
certificates. E-mail us for information on this rather costly
apparatus ($8900.00 without computer).
Test indicators can also be calibrated on a surface plate using
certified gage blocks. The indicator is securely fastened to a stand
and the contact point is brought in contact with a gage block of a
given size. The contact point must be parallel with the surface of
the block for most manufacturers. Interapid test indicators are an
exception and should be at a 12-degree angle, approximately. The
gage block can then be removed and replaced a number of times to
check for repeatability. Be certain that discrepancies in
repeatability are not due to poorly tightened clamps, flimsy stands
or other factors. Usually one quarter of a graduation repeatability
is allowable, but check with the manufacturer's calibration specs
for your particular model.
Errors in repeatability indicate a need for cleaning and, possibly,
repair. Do not attempt this without experience.
Accuracy in travel is checked by replacing the gage block with one
larger. Very small intervals are required. Ideally you'd want to
check the travel at every half revolution, or better. During this
procedure be certain that the gage blocks are properly wrung to each
other and to the surface. In general, accuracy should not vary more
than one graduation per dial revolution on .0005" indicators.
If an incremental error occurs - one which increases regularly over
the entire travel - then the contact point is of the wrong length or
the angle of the point in regard to the surface is incorrect. You
should verify that the correct length point is being used.
Furthermore, you can make small adjustments by changing the contact
point angle. Make repeated calibration attempts with varying angles
until you find one which gives correct results. Obviously, it will
now be necessary to recreate this same angle when the indicator is
used in actual test situations. Some indicators (Girodtast, for
example) allow you to make small adjustments in length with a set
screw.
One final method requires a certified height master. This takes the
place of gage blocks. The one we use has an accuracy of
.00002". The test indicator is firmly fastened to a test stand
and the contact point is positioned (at the proper angle) over one
of the height master's test surfaces. Comparison readings are now
taken at half-revolution intervals - or better - in both
directions.
About the cosine error: for test indicators excluding Interapid
models. If the contact point can not be kept parallel to the work
surface then you will have to make a mathematical adjustment to the
dial reading.
Contact point angle correction factor
10° reading times 0.98
15° reading times 0.97
20° reading times 0.94
30° reading times 0.87
40° reading times 0.77
50° reading times 0.64
60° reading times 0.50
From this chart you will notice that a contact point held at a
60-degree angle results in one-half the dial reading. Once you
determine the angle, simply multiply the dial reading by the
corresponding correction factor.
For example, an indicator reading of .0085" at an angle of
30-degrees is equivalent to
.0085" x .87 = .0074"