You should check
height gages
like you check vernier calipers—with the addition of checking the
base-to–column squareness. Base-to–column squareness must be
checked from a surface plate, so surface plate errors must also be
checked. Therefore, height gage inspection and calibration should
be done by the gage inspection department.
You can perform several tests on your own, however, that show
whether the instrument needs to be calibrated. Regular visual
inspection can reveal burrs, which can be removed, and nicks that
may indicate sever abuse or damage. In addition, the base should
not rock on a flat surface, which you can determine by applying
pressure at various places. A truly flat base will “wring�?to a
fine surface plate and you will feel drag or suction when you
slide the instrument.
The best way to test accuracy and precision is to make repeated
readings of gage block height at various points throughout the
range. Again, it is not enough to repeat the measurements. If ,
after you have corrected the zero setting by adjusting the vernier
plate, the precision (dispersion of repeated readings) or the
accuracy (repeating readings of the gage block height ) is more
than the discrimination of the instrument (20Чm /0.001 in.), you
should return the instrument to the manufacture for repair. You
should always inspect your accessories as carefully as you inspect
the instrument itself.